The first reported case of a pacemaker explosion during cremation was in 19765. The body of a 70-year-old man was cremated at 800 °C. After 5 minutes, four explosions occurred in rapid succession with a final explosion a few minutes later. In the wall of the cremator was a finger-sized hole half an inch deep. Among the cremated remains, there were five discs resembling the ends of rifle cartridges, a short length of wire and a metal plate. The device was identified as a zinc/mercuric oxide pacemaker. These pacemakers explode on cremation because of the rapid formation of hydrogen gas which bursts the pacemaker casing6.
Today, most pacemakers have a lithium/iodine-polyvinylpyridine (PVP) batterybecause of its greater longevity, the smaller cell size and predictable depletion7. Other pacemaker power sources have included zinc/mercuric oxide, nickel cadmium and plutonium-2388. Battery development plays a key role in pacemaker design because small-volume high-energy power sources are required. This requirement, the increasing pacemaker prevalence and the greater number of cremations in the UK (from 3 in 1885 to 44 169 in 1999, now accounting for over 70% of funerals)9 may result in increasingly powerful and frequent explosions in crematoria. We aimed to determine the frequency and consequences of pacemaker explosions in crematoria, and investigated the current procedures for preventing them.
| METHODS |
|---|
|
|
|---|
| RESULTS |
|---|
|
|
|---|
|
Of the 71 crematoria that had reported distress or damage as a result of pacemaker explosions, the commonest consequences were the noise of the explosion and damage to the cremator doors and brickwork32 (45%) and 30 (42%) crematoria, respectively. The cremator was damaged beyond repair in 3% of cases and in one case the explosions caused injury to staff. The remains of pacemakers were found after 15% of explosions.
The procedure for checking that pacemakers were removed before cremation was as follows. 99% of crematoria staff check the cremation forms to ensure that the doctor signing the form has confirmed that there is no pacemaker in the body. 54% of crematoria staff also discuss the case with the funeral director to check that the pacemaker has been removed from the body. Only 9% of crematoria staff ask the relatives whether a pacemaker is present. 16% make other enquiries (e.g. speaking to the coroner's office, mortuary staff, hospital, or medical referee or placing a reminder slip with the cremation form for the attention of the attending doctor).
Most crematoria staff believe that checking the cremation form is the best method of ensuring that pacemakers have been removed before cremation. However, only 5% of them knew about implantable cardiac defibrillators and their explosive potential (one centre reported a large explosion caused by the cremation of a body containing an implantable cardiac defibrillator). 54% believe that a hand-held metal detector might help identify pacemakers and other implantable devices that could explode.
| DISCUSSION |
|---|
|
|
|---|
Today, most pacemakers are driven by the lithium/iodine-PVP energy source. At room temperature these devices are benign. However, during cremation, when temperatures reach 1300 °C (2400 °F) for 90 minutes, iodine forms a gas that rapidly expands, causing the pacemaker casing to burst. A chemical reaction also causes an explosion: at 180.5 °C lithium melts and reacts with the gaseous iodine to release in less than 1 second the energy which would be expended over several years (about 64 kcal/mol).
Pacemakers now in the design stage will be potentially more explosive and also more difficult to detect post mortem: both manufacturers and patients favour smaller pacemakers that have greater energy. Solid cathode, liquid electrolyte systems such as the lithium/carbon monofluoride and lithium/manganese dioxide pacemakers have greater gravimetric energy density (watt h/mm3) and are therefore likely to be future cardiac pacemaker power sources10.
Cremation forms must be completed by medical staff to prevent the inappropriate cremation of pacemakers. Our study demonstrates that most crematoria staff rely on a completed and accurate cremation form to ensure that pacemakers are not present in the body. Since it is against the code of practice of crematoria staff to open sealed coffins, they depend on others to provide accurate information. Indeed, many crematoria staff discuss the issue of pacemakers with funeral directors, who are able to inspect the body in an attempt to prevent the cremation of pacemakers.
As a result of the first reported incident, in 19765, two supplementary questions were added to form B of the Cremation Act certificate. They remain in use and ask (a) Has a pacemaker or any radioactive material been inserted in the deceased (yes or no)?; (b) If so, has it been removed (yes or no)? If (b) is answered in the negative, the medical referee may, under Regulation 12 of the Cremation Regulations (1930), prevent cremation4. Similar questions were advised for the coroner's form E, but have not been implemented. It has been argued that the cremation form should be abolished and the medical certificate amended to include questions about pacemakers as well as suspicion about the cause of death11.
Despite the statutory questions on the cremation form B, we have found that some pacemakers remain in the body during cremation. There are several possible explanations for this. First, the doctor signing the form may mistakenly report that there is no pacemaker in the body or the form may be incomplete. As a result, pacemakers are left in the body. There is little doubt that this occurs: only 41% of forms presented to a crematorium were completed sufficiently for the staff to proceed without further enquiry12. In another survey, 6% of cremation forms had incorrect or incomplete details of pacemaker or radioactive implants, with over 25% of forms being unsatisfactorily completed13. Secondly, the coroner's form E does not contain the statutory pacemaker question. With the coroner's certificate now accounting for about a third of disposals16, it may be that some pacemakers remain in situ. Thirdly, it is possible that only one pacemaker is removed from a body that contains two or more devices. Finally, pacemakers can migrate in the subcutaneous tissue, making detection by palpation difficult.
Pacemaker explosions are potentially costly for the crematorium owner and may lead to legal proceedings against funeral directors, physicians and health authorities to recover losses. A hand-held metal detector (such as those used at airports) might help detect pacemakers in the mortuary. Our study shows that many of those surveyed believe that such a device could be useful. This warrants further study. Other metallic devices (such as syringe drivers and prostheses) would also be detected, but this could be verified with reference to the deceased's medical case-notes.
Our study is a retrospective analysis of returned questionnaires. Probably there is under-reporting of the pacemaker explosions. Staff may not wish to mention these events and their recall may not be accurate. Such a discrepancy is apparent in our results because 47% of staff who responded reported personal experience of pacemaker explosions at some point in their career, but only 41% of staff estimated no events occurring in general. To our knowledge, there is no routine collection of data on explosions in UK crematoria. If this were done, we would have an accurate record of the frequency and consequences of explosions.
Our study shows how much crematoria staff rely on accurate and complete cremation forms. Physicians entitled to sign forms B and C must be aware of the legal and financial responsibilities of certification and the consequences when pacemakers are heated. Simple steps such as the duplication of the two statutory questions on form B to the coroner's form E, along with a question about implantable devices on the relatives' form A, might help reduce the number of explosions in crematoria. In addition, the routine placement of a warning label on the front of all the clinical notes of people with pacemakers and other implantable devices would alert the physician to the presence of potentially explosive material in the body.