| SUMMARY |
|---|
|
|
|---|
53 devices measured arm BP, 21 measured forearm BP. The agreement between the mercury sphygmomanometer and the automated devices was poor, with average differences of 9.5 mmHg for systolic and 9.4 mmHg for diastolic and no clear advantage for either site of measurement. As a means of screening for BP >140/90 mmHg the sensitivity of the automated devices was 81% and the specificity was 80%. There were large variations in how often and under what circumstances the devices had been used. One-fifth of the devices had been acquired on medical advice but only 11% of the participants were aware of the three important conditions for operating such devices.
Discussion of automated devices, their role and proper use, should now be part of routine hypertensive care.
| INTRODUCTION |
|---|
|
|
|---|
The accuracy of home BP monitoring depends firstly on the machine's intrinsic reliability but also on the way the machine is used. For determining adequacy of BP control, readings should be taken at the same time of the day, preferably in the early morning and occasionally between medications.10,11 Correct technique is critical, and the measurement should be repeated after 5 minutes or more.12
In clinical practice we observed that automated BP devices are very popular in Hong Kong and their use is sometimes anomalous. We therefore conducted a small study to assess the use of such devices in Chinese hypertensive patients and how they reacted to abnormal readings.
| METHODS |
|---|
|
|
|---|
The clinic record systems were used to generate a list of patients aged between 40 and 70 years who had visited the clinic with the diagnoses of raised BP, hypertension with or without end-organ diseases (ICPC-2 codes K85, 86 and 87), between 31 January 2002 and 1 February 2003. A research assistant phoned all these patients, asking if they had an automated home BP device and had been using it for the last three months. Patients who could not be contacted by phone on three separate occasions were classified as lost to contact. To detect a difference of 10% between the two methods at a two-tailed alpha of 0.05 and power of 80% the estimated sample size was 72.
The patients were requested to attend the health centre on two occasions 714 days apart and to avoid smoking, alcohol, eating, bladder distension and exercise for at least 30 minutes before the visits. During the first visit, the research assistant explained the study and asked patients to sign a consent form. Then, during a 10-minute interview, they filled in a questionnaire on knowledge and use of the automated BP device. The participants then measured their BP with their own automated device without help; and about 15 minutes afterwards the research assistant repeated the measurement with a mercury sphygmomanometer. Both procedures were repeated 10 minutes later. During the second visit, the same method and sequence were employed. Therefore, a total of eight BP readings were collected for each participant, four from the automated and mercury sphygmomanometers, and four from the first and second visits. Comparisons between the measurements were made by use of Bland-Altman plots.
| RESULTS |
|---|
|
|
|---|
|
The average absolute difference between measurements with the patient's BP device and the mercury sphygmomanometer for systolic BP (SBP) was 9.5 mmHg (P=0.40) and for diastolic BP (DBP) 9.4 mmHg (P=0.08). For SBP, 39% of readings had an absolute difference below 5 mmHg, 62% of them below 10 mmHg and 81% below 15 mmHg whereas, for DBP, the respective percentages were 36%, 64% and 84%. Figures 1 and 2 illustrate the poor agreement between automated devices and mercury sphygmomanometerconcordance correlation coefficient for SBP 0.84, for DBP 0.52.
|
|
We examined the potential value of the devices for detecting hypertension as defined by the World Health Organization, 4140/90 mmHg. When compared to the mercury standard, the sensitivity of all BP devices used by our sample was 81% (17/21) and the specificity was 80% (218/273). On multivariate logistic regression neither price (5£42.7; P=0.08) nor site of measurement (forearm/arm; P=0.23) was shown to determine whether a machine was 'good' as reflected by the European Society of Hypertension's grade C or above.14
85% of patients said they had had no training on how to operate their machine and only 63% had read the user's manual. About half knew that the position of the cuff should be level with the heart and that they should rest for 5 minutes or more before taking the BP and one-fifth appreciated the importance of appropriate cuff size; although 20% of patients had acquired the device on medical recommendation, only 11% were knowledgeable on all three points.
The frequency with which patients measured their BP varied widelyfrom more than once a day (4 patients) to less than once a month (16). Two-thirds used the machine when they had symptoms (e.g., headache, dizziness) and one-third for monitoring BP control. Only 11 patients had consulted their doctor after finding a raised BP, and the medication had been changed in 3 of these.
| DISCUSSION |
|---|
|
|
|---|
Ownership of automated BP devices is widespread, and we suspect that experience in other developed countries would not differ greatly from ours in Hong Kong. Inaccuracy, through intrinsic properties or incorrect use, can cause harm. False positives can result in unnecessary anxiety; false negatives can provide misleading reassurance and reduce the incentive to change risk behaviours. On the other hand, home measurements do allow patients to take some responsibility for their own care, and BP control is likely to be best in those who are well informed about their condition. We suggest that, whenever hypertension is diagnosed, the advantages and limitations of self-monitoring should come into the discussion. Patients who decide to use home monitors should be trained in the techniques and advised on what to do when their readings are abnormal.